Line 7: Line 7:
 
Tref 20.4C
 
Tref 20.4C
  
 +
Latest Factory Cal
 +
pb275 1-08-2016
 +
Tcal 18.1
  
 
HS-6 UC Calibration
 
HS-6 UC Calibration
 
09-2013 pb252 on should use this cal
 
09-2013 pb252 on should use this cal
 
*Check gain values in calibration file for sigmacorrect script
 
*Check gain values in calibration file for sigmacorrect script
09/2013 kexp [.15 .148 .15 .15 .152 .150]
+
09/2013 kexp [.150 .148 .150 .150 .152 .150]
 +
 
 +
10/2014 kexp [.142 .141 .142 .143 .145 .141]
 +
 
 +
05/2015 kexp [.137 .137 .136 .139 .143 .136]

Latest revision as of 09:53, 23 May 2016

AC9 Factory cal 10/2012 pb243 and on should have this cal.

  • update goac9
  • update ac9cal.txt
  • check aw and bw data with cal CD

Tref 20.4C

Latest Factory Cal pb275 1-08-2016 Tcal 18.1

HS-6 UC Calibration 09-2013 pb252 on should use this cal

  • Check gain values in calibration file for sigmacorrect script

09/2013 kexp [.150 .148 .150 .150 .152 .150]

10/2014 kexp [.142 .141 .142 .143 .145 .141]

05/2015 kexp [.137 .137 .136 .139 .143 .136]