Difference between revisions of "Calibration process procedures"
Line 1: | Line 1: | ||
'''After calibration processing of calibration files''' | '''After calibration processing of calibration files''' | ||
− | As of October 2010 calibrations of been reprocessed with new ' | + | As of October 2010 calibrations of been reprocessed with new 'Si<sub>t</sub>' values. The processing directory for AC9 calibrations is ''/pb/REPROCESS_2010/AC9/cals_reprocess/ '' instrument number(184) ''/QWATER/'' year/date folder. |
To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in. ac9cal.m sources ac9cal_def.par in the ''cals_reprocess/LIST'' folder. | To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in. ac9cal.m sources ac9cal_def.par in the ''cals_reprocess/LIST'' folder. |
Revision as of 11:22, 24 November 2010
After calibration processing of calibration files
As of October 2010 calibrations of been reprocessed with new 'Sit' values. The processing directory for AC9 calibrations is /pb/REPROCESS_2010/AC9/cals_reprocess/ instrument number(184) /QWATER/ year/date folder.
To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in. ac9cal.m sources ac9cal_def.par in the cals_reprocess/LIST folder.
Reviewing calibrations
After processing, it is good to compare the latest calibration processed values with previous ones. For this, data from the _ave.txt files(which should be linked by you from the .txt files in the year/date folder) is added to the qcal_a_yy_yy.txt and qcal_c_yy_yy.txt files. ac9cal.m is restarted and the qcal file is now the *yy_yy.txt folder. Hit the last button ts_cal to view the whole time series.
back to Main Page