'''After calibration processing of calibration files'''  
 
'''After calibration processing of calibration files'''  
  
As of October 2010 calibrations of been reprocessed with new 'Si<sub>t</sub>' values.  The processing directory for AC9 calibrations is ''/pb/REPROCESS_2010/AC9/cals_reprocess/ ''  instrument number(184) ''/QWATER/'' year/date folder.
+
<big> AC-9<big>
 +
 
 +
As of October 2010, AC-9 calibrations have been reprocessed with new 'Si<sub>t</sub>' values from Wetlabs AC meter protocol document, rev.P.  The processing directory for AC9 calibrations is ''/pb/REPROCESS_2010/AC9/cals_reprocess/ ''  instrument number(184) ''/QWATER/'' year/date folder.
  
 
To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in.  ac9cal.m sources ac9cal_def.par in the ''cals_reprocess/LIST'' folder.
 
To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in.  ac9cal.m sources ac9cal_def.par in the ''cals_reprocess/LIST'' folder.

Revision as of 16:03, 11 March 2011

After calibration processing of calibration files

AC-9

As of October 2010, AC-9 calibrations have been reprocessed with new 'Sit' values from Wetlabs AC meter protocol document, rev.P. The processing directory for AC9 calibrations is /pb/REPROCESS_2010/AC9/cals_reprocess/ instrument number(184) /QWATER/ year/date folder.

To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in. ac9cal.m sources ac9cal_def.par in the cals_reprocess/LIST folder.

Reviewing calibrations

After processing, it is good to compare the latest calibration processed values with previous ones. For this, data from the _ave.txt files(which should be linked by you from the .txt files in the year/date folder) is added to the qcal_a_yy_yy.txt and qcal_c_yy_yy.txt files. ac9cal.m is restarted and the qcal file is now the *yy_yy.txt folder. Hit the last button ts_cal to view the whole time series.


back to Main Page