Difference between revisions of "Calibration process procedures"

From Pnb
Jump to navigation Jump to search
Line 3: Line 3:
 
<big>AC-9</big>
 
<big>AC-9</big>
  
As of October 2010, AC-9 calibrations have been reprocessed with new 'Si<sub>t</sub>' values from Wetlabs AC meter protocol document, rev.P.  The processing directory for AC9 calibrations is ''/pb/REPROCESS_2010/AC9/cals_reprocess/ ''  instrument number(184) ''/QWATER/'' year/date folder.
+
As of October 2012, AC-9 calibrations have been reprocessed with new 'Si<sub>t</sub>' values from Wetlabs AC meter protocol document, rev.Q.  The processing directory for AC9 calibrations is ''/pb/AC9/cals/ ''  instrument number(184) ''/QWATER/'' year/date folder
 +
 
 +
Each time the instrument is sent back for a factory calibration, the new disc should be checked for updated temp and sal correction coefficients and calibration history should be updated in the ac9cal.txt inside of the ''/cals/0184/DeviceFiles/'' directory.  Note that if coefficients have changed, goac9.par will need to be updated in the fallowing cruise processing.
  
 
To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in.  ac9cal.m sources ac9cal_def.par in the ''cals_reprocess/LIST'' folder.
 
To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in.  ac9cal.m sources ac9cal_def.par in the ''cals_reprocess/LIST'' folder.

Revision as of 16:36, 24 October 2012

After calibration processing of calibration files

AC-9

As of October 2012, AC-9 calibrations have been reprocessed with new 'Sit' values from Wetlabs AC meter protocol document, rev.Q. The processing directory for AC9 calibrations is /pb/AC9/cals/ instrument number(184) /QWATER/ year/date folder.

Each time the instrument is sent back for a factory calibration, the new disc should be checked for updated temp and sal correction coefficients and calibration history should be updated in the ac9cal.txt inside of the /cals/0184/DeviceFiles/ directory. Note that if coefficients have changed, goac9.par will need to be updated in the fallowing cruise processing.

To process the calibration files, the device number _watertemp.txt file must be updated with the latest temperature value and .dat files must be present in the folder you wish to process in. ac9cal.m sources ac9cal_def.par in the cals_reprocess/LIST folder.

Reviewing calibrations

After processing, it is good to compare the latest calibration processed values with previous ones. For this, data from the _ave.txt files(which should be linked by you from the .txt files in the year/date folder) is added to the qcal_a_yy_yy.txt and qcal_c_yy_yy.txt files. ac9cal.m is restarted and the qcal file is now the *yy_yy.txt folder. Hit the last button ts_cal to view the whole time series.


HS-6